User’s Guide
Model 445814Humidity Alert
Humidity Indicator with Dew Point and Programmable Hi/Lo %RH Alarms
Introduction
Congratulations onyour purchase of Extech’s Humidity Alert Hygro-Thermometer.
Humidity alarms warn when the%RH exceeds a pre-set HI orLO limit. It also
features Humidity and Temperature calibration adjustments. The unit can be wall-
mounted or placed on a flat surface usingthe fold-out stand. This professional
meter, with propercare,will provide years ofsafe reliable service.
Meter Description
1.Temperature-Humidity Sensor
2.%RH Display
3.Dew Point Display
4.Temperature Display
5.%RH LED alarm
6.Max/Min push button
7.Reset push button
8.Audible Alarm ON/Off switch
9.Tilt Stand
10.Humidity calibration adjust
11.Temperatureunits switch (ºF/ºC)
12.Temperature calibration adjust
13.Wall hanger
14.Battery compartment
Cautions
This device is not atoyand must not reach children’s hands. It contains
hazardous objects as well as small parts that thechildren couldswallow. In
case a child swallows any of them, pleasecontact a physician immediately
Do notleave batteries and packing material lying around unattended; they
can be dangerous forchildren if they use them as toys
In case the device isgoing to be unused for an extended period of time,
remove the batteries to prevent themfrom training
Expired or damaged batteries can cause cauterization on contact with the
skin. Always, therefore, use suitable handgloves in such cases
See that the batteriesare not short-circuited. Do not throw batteriesinto the
fire.
Copyright © 2005 Extech Instruments Corporation. All rights reserved including
the right of reproduction in whole or in partin any form.
www.extech.com
V3.1 10/05
Operation
Preparation foruse
Open the battery compartment by pushing the cover on the rearof the
instrument downward as indicated by thearrow. Remove thebattery insulating
strip and the protective film cover on the LCD display.
Temperature Units
Select ºF or ºC unitsusing the rearselection switch.
Audible alarm
Selectorusing the rear selection switch to turn the audiblealarmon or
off.
Maximum /Minimum display
1.Press the MAX/MINbutton. The highest value measured since the Reset
button was last pressed appears on the display.
2.Press the MAX/MINbutton again to display the lowest value measured since
the Reset button was last pressed.
3.Press the MAX/MINbutton again to return to normaldisplay operation.
Resetting (clearing)the MAX/MINmemory
1.Press the MAX/MINbutton to enter the MAX/MIN mode.
2.Press and hold the RESET button for1 second to clearthememoryand
start recording new max/min values.
Setting the %RH AlarmLimits.
1.SimultaneouslyPress and HOLDthe MAX/MIN and RESET buttonuntil “HI”
and the stored HI%RH limitappears flashing.
2.Press the RESET button to set the Hi%RH limit. Hold thebutton in for rapid
adjustment.
3.Press the Max/Min button. “LO”and the stored LO %RH limit will appear
flashing
4.Press the RESET button to set the LO%RH limit.
5.Press the Max/Min button to turn on or offthe alarm. The alarmiconwill
appear in the display when the alarmis active.
6.Press and hold the RESET button to save the settingsand exit the mode.
Alert
If %RH exceeds theHI limit or goes below the LO limit thealert will be
activated. The alert consist of two flashing ‘ALERT” icons in the upper display,
a flashing red light (for60 seconds) and an audible buzzer (for60 seconds)
Calibration
RH Calibration
1.Insert meter’s sensor intoa humidity chamber.The reference should be >70%
RH.
2.Check the reading after1 hour.
3.Adjust the RH calibrationpot in ½turn increments, waiting for the display to
update after each adjustment, until the reading is within the accuracy
specification.
Temperature Calibration
1.Place the sensor in astabilized environment ofapproximately21°C.
2.Check the reading after1 hour.
3.Adjust the temperature calibration pot in ½turn increments, waitingfor the
display to update aftereach adjustment, untilthe reading is within the
accuracy specification.
Battery Replacement
Open the battery compartment by sliding the battery cover on the rear of the
instrument downward as indicated by thearrow. Replace the 1.5V ‘AA’batteries
and replace the cover.
You, as the end user,are legally bound (Battery ordinance) to
return allused batteriesand accumulators;disposal in the
household garbage is prohibited!
You can hand over yourused batteries / accumulators,
gratuitously, at the collection points forourbranches in your
community or wherever batteries / accumulators are sold!
Disposal: Follow the valid legalstipulations in respect of the
disposal of the device atthe end of its lifecycle
Specications
Range Accuracy
Temperature-10 to 60
oC
14 to 140
oF
+ 1.0oC within -10oC to 50oC
+ 1.8oF within 14oF to 122oF
Humidity10% to 99%+
4% RH within 25%to 85% RH and 0to
50oC or 32 to 122oF
Dew Point-36 to 60
oC
-32.8to 140
oF
+/-2oC.@ 20
oC
+/-3.6oF @ 68oF
Battery(2) 1.5V 'AAA' cell
Weight6 oz (169g)
DimensionsDisplay: 4.3x3.9x0.8” (109x99x20mm)
Note: Ifsubjected to anelectrostatic discharge themeter may
malfunction and require the user toperform a reset.
Specyfikacje produktu
| Marka: | Extech |
| Kategoria: | Sprzęt pomiarowy |
| Model: | 445814 |
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Extech 445814, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje Sprzęt pomiarowy Extech
18 Marca 2025
30 Stycznia 2025
8 Stycznia 2025
8 Stycznia 2025
8 Października 2024
8 Października 2024
8 Października 2024
8 Października 2024
8 Października 2024
8 Października 2024
Instrukcje Sprzęt pomiarowy
Najnowsze instrukcje dla Sprzęt pomiarowy
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025