PN 2811813 April 2007 © 2007 Fluke Corporation. All rights reserved. Printed in U.S.A. 1
Fluke-700LTP-1
Low-Pressure Test Pump
Instruction Sheet
Introduction
The Fluke-700LTP-1 Low-Pressure Test Pump is a
portable, dual source of vacuum and pressure. Each pump
incorporates a vacuum and pressure selector and fine
adjustment control. The pump has the following
specifications:
• Output pressure: 0 to 100 psi / 0 to 6.9 bar
• Output vacuum: 0 to -13 psi / 0 to -900 mbar
• Exposed Materials: Aluminum, silicon, neoprene,
stainless steel, Buna-N
• Adjustment: Fine volumetric pressure and vacuum
adjuster
• Dimensions: 150 mm x 100 mm
• Weight: 226 grams or .5 lbs (pump only)
Box Contents
• Fluke 700LTP-1 Low-pressure Test Pump
• 1/8 in NPT male/female tee
• (2) 1 m hoses
• (4) 1/8 in NPT male quick connects
• (2) 1/8 NPT female to 1/4 in BSP female
• Seal Kit
• Instruction Sheet
Contacting Fluke
To contact Fluke or for service, call one of the following
telephone numbers:
USA: 1-888-44-FLUKE (1-888-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.
1
2
8
3
5
6
7
4
eus01f.eps
A
Pressure/vacuum release valve- Turn valve counter-
clockwise for vacuum, clockwise for pressure and
centered for release.
B Fine adjustment control- Turn in or out to increase or
decrease pressure accordingly.
C Master and test port tee
D Quick-fit connectors
E Hose
F 1/8 in NPT female to 1/4 in BSP female fitting
G
Pump Handle. If the pump has not been used for a
while, there may be some resistance on the first
stroke.
H Cleanout ports (2)- See cleanout instructions
Figure 1. Features
Guidelines for Use
1. Connect a pressure calibrator or pressure module to
the quick-fit connector (item D) at the end of one of the
flexible hoses (item E).
2. Connect the unit under test to the quick-fit connector
(item D) at the end of the second flexible hose (item
E F), choosing correct adapters (item ) and seals.
3. Turn the fine adjustment control to mid-range.
4. Set control valve to pressure.
5. Operate handle (item G) until the pressure is close to
the required value.
Specyfikacje produktu
Marka: | Fluke |
Kategoria: | Przepych |
Model: | 700LTP-1 |
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Fluke 700LTP-1, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje Przepych Fluke
24 Września 2024
21 Września 2024
14 Września 2024
21 Sierpnia 2024
21 Sierpnia 2024
20 Sierpnia 2024
19 Sierpnia 2024
6 Sierpnia 2024
Instrukcje Przepych
- Blackburn
- Blumfeldt
- Lavorwash
- Oase
- Fieldmann
- Simpson
- SilverCrest
- Polaris
- Makita
- Milwaukee
- JANDY
- DAB
- Esotec
- Pontec
- Barwig
Najnowsze instrukcje dla Przepych
5 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025
1 Kwietnia 2025