Installation Manual
Type
Medical Switching Power Supply
(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)
Installation
(1) Enclosed Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Enclosed units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Enclosed units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.
(2) Modular Type
(a) Before any installation or maintenance work, please disconnect your system from the utility.
Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient
temperature may increase the internal component temperature and will require a de-rating in
output current. Please refer to the specification sheets to receive the optimum mounting
position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance
must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of
Modular units. Please refer to case drawing on specifications to receive the maximum length of
mounting screw.
(e) Recommended wires of Modular units are shown as below.
AWG 18 16 14 12 10 8
Rated Current of
Equipment (Amp) 6A 6-10A 10-16A 16-25A 25-32A 32-40A
Cross-section of
Lead(mm2 ) 0.75 1.00 1.5 2.5 4 6
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more
strips should not exceed 80% of current rating.
Make sure that all strands of each stranded wire enter the terminal connection and the screw
terminals are securely fixed to prevent poor contact.
(f) For other information about the products, please refer to www.meanwell.com for details.


Specyfikacje produktu

Marka: Mean Well
Kategoria: без категории
Model: RPS-160-12

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Mean Well RPS-160-12, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje без категории Mean Well

Instrukcje без категории

Najnowsze instrukcje dla без категории

Technaxx

Technaxx TX-182 Instrukcja

23 Września 2024
Holman

Holman BTX1 Instrukcja

22 Września 2024
Gravity

Gravity Vari-G 7 Instrukcja

22 Września 2024
Ridgid

Ridgid RE 60 Instrukcja

22 Września 2024
ATen

ATen VE-RMK1U Instrukcja

22 Września 2024
LG

LG WF-T7016PP Instrukcja

21 Września 2024