Instruction Manual
Model: ThermoCal
Professional Testing Device
for Thermometer
www.rossmax.com
OBM_IB_ThermoCal_ver1411
Manufacturer: Rossmax Swiss GmbH,
Tramstrasse 16, CH-9442 Berneck, Switzerland
2. Heating indicator of the water bath does not light up after the calibration temperature is set
Possible causesActions to take
Water temperature is higher than the set
temperature Cool down the water in the bath
Defective heating indicatorContact Rossmax for service
Defective temperature controllerContact Rossmax for service
Set temperature is too lowSet temperature should be at least 5°C higher
than operating temperature
3. Heating indicator of water bath lights up, whereas the water temperature remains low
Possible causesActions to take
Defective heater circuitContact Rossmax for service
Defective temperature controllerContact Rossmax for service
4. Water temperature continues to rise and exceed the set temperature
Possible causesActions to take
Defective temperature controllerContact Rossmax for service
5. Large uctuation in water temperature
Possible causesActions to take
Water temperature is not stable yetWait for temperature to stabilize
Set temperature is too close to ambient
temperature Lower the ambient temperature
Defective temperature controllerContact Rossmax for service
Defective water heaterContact Rossmax for service
6. Power and power indicator are both on, whereas the water circulating pump does not start
Possible causesActions to take
Foreign object clogs inside the pumpRemove the foreign object
Loose screw used to secure the propellerTighten the screw
Defective motorContact Rossmax for service
7. Noise generated by the stirrer motor
Possible causesActions to take
Bearing ball oxidationReplace bearing
Bearing agingReplace bearing
Trouble shooting
1. Power is switched on but the power indicator does not light up
Possible causesActions to take
No power inputCheck the power source
Loose plugInsert the plug into the socket securely
Device is powered and running but the power
indicator does not light upReplace the power switch
Power wire broke or loose connectionReplace the power wire(s)
Tripped fuseCheck the fuse and replace it if needed
WARNING: The symbol on this product means that it’s an
electronic product and following the European directive
2012/19/EU the electronic products have to be disposed
on your local recycling centre for safe treatment.
Specyfikacje produktu
| Marka: | Rossmax |
| Kategoria: | termometr |
| Model: | ThermoCal - Professional Testing Device for Thermometer |
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Rossmax ThermoCal - Professional Testing Device for Thermometer, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje termometr Rossmax
4 Października 2024
3 Października 2024
3 Października 2024
13 Września 2024
7 Września 2024
3 Maja 2024
3 Maja 2024
Instrukcje termometr
Najnowsze instrukcje dla termometr
8 Kwietnia 2025
7 Kwietnia 2025
6 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
3 Kwietnia 2025
2 Kwietnia 2025
2 Kwietnia 2025
2 Kwietnia 2025
1 Kwietnia 2025