3 - -GB_V1.5 9/19 80400 en
User Manual
Resistance Decade Box
Model 380 0 40
Introduction
Congratulations on your purchase of the Extech 380400 Resistance Decade Box. This device
oers 7 decades of resistance ranges from 1 to over 11MOhms (in 1Ohm steps). Slide switches
allow easy addition and subtraction of resistance values. Binding posts (3) oer simp and le
secure connections. Careful use of this decade box will provide years of reliable service.
Specications
Resistance ranges 1 to 11,111,110 Ohms in 1 Ohm steps
Internal Resistance 0.3 Ohms
Power 0.3W Resistors
Connection Three (3) binding posts
Accuracy 1% Reading) + 0.3 Ohms ±(
Operating conditions Temperature: 32 to 122oF (0 to 50oC) / Humidity: 80%RH <
Dimensions/Weight 5.79 x 4.61 x 1.3" (14.7 x 11.7 x 3.3 Approx. 0.69 lbs. (312 cm) / g)
Operation
Binding Post Connections
The binding posts can be used for connections in several ways:
1. A banana plug can be inserted directly into the posts.
2. Bare wire can be threaded through the post after it has been unscrewed. Once the bare wire is threaded, tighten the posts as
necessary.
3. Alligator clips can be used but use caution not to strip the post threads or plastic post housing.
The resistance output is available on the RED and the BLACK binding posts. The WHITE post is case ground and is typically not
used. Connect the positive lead of the device under test to the RED post. Connect the negative lead to the BLACK post. Use the
WHITE grounding post only if the device under test will be grounded to the 380400 case.
Range Selection
The front panel switches are used to select the resistance that will be available on the RED and BLACK terminals. When a switch 28
is set to the IN position, the value printed above the switch is added to the total resistance available at the posts. When the switch is
set to OUT it is excluded from the total resistance. If all of the switches are set to OUT, the total output resistance zero (+ 0.3Ohms is
internal resistance approx.). –
For example, if the desired output value is 10.5KOhms, set the following switches to the IN position: 10K, 400Ohms and 100Ohms , .
Testing
This device can be used to verify the calibration integrity of MultiMeters, LCR meters, Calibrators, etc. Connect as described above in
the Binding Post Connection section. Set the resistance switches to output the desired resistance. Ensure that the voltage supplied
by the device under test does not overload the 0.3W internal resistor power rating. The device under test should read the value of
resistance selected on the tester. If it does not, the device under test may need calibration, adjustment, or repair.
Copyright - FLIR Systems, Inc. ©2013 2019
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com
Specyfikacje produktu
Marka: | Extech |
Kategoria: | Niesklasyfikowane |
Model: | 380400 |
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Extech 380400, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje Niesklasyfikowane Extech
8 Stycznia 2025
3 Października 2024
3 Października 2024
3 Października 2024
29 Września 2024
29 Września 2024
27 Września 2024
19 Września 2024
18 Września 2024
13 Września 2024
Instrukcje Niesklasyfikowane
- Niesklasyfikowane BWT
- Niesklasyfikowane Maginon
- Niesklasyfikowane Onkyo
- Niesklasyfikowane Canyon
- Niesklasyfikowane Baby Brezza
- Niesklasyfikowane Potenza
- Niesklasyfikowane Auto Joe
- Niesklasyfikowane Sanitas
- Niesklasyfikowane Global Water
- Niesklasyfikowane Roland
- Niesklasyfikowane Vimar
- Niesklasyfikowane Seville Classics
- Niesklasyfikowane Omega
- Niesklasyfikowane Cypress
- Niesklasyfikowane Flex
Najnowsze instrukcje dla Niesklasyfikowane
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025
29 Stycznia 2025