Instrukcja obsługi Kyoritsu 2434


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Kyoritsu 2434 (2 stron) w kategorii multimetr. Ta instrukcja była pomocna dla 27 osób i została oceniona przez 14 użytkowników na średnio 4.7 gwiazdek

Strona 1/2
INSTRUCTIONMANUAL
KEW SNAP
 
2434
KEW SNAP Series
LEAKAGECURRENTTESTER
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
1. SAFETY WARNINGS
This instrument has been designed and tested according to IEC
Publication 61010: Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety rules
which must be observed by the user toensuresafeoperation of the
instrumentandtoretainitinsafecondition.Therefore,readthroughthese
operatinginstructionsbeforestartingusingtheinstrument.
#
WARNING
Read through and understand instructions contained in this manual
beforestartingusingtheinstrument.
Saveandkeepthemanualathandtoenablequickreferencewhenever
necessary.
Besuretouse the instrumentonlyinitsintended applications andto
followmeasurementproceduresdescribedinthemanual.
Be sure to understand and follow all safety instructions contained in
themanual.
Notfollowingtheaboveinstructionsmaycauseinjury,instrumentdamage
and/ordamagetoequipmentundertest.
Theinstrumentistobeusedonlyinitsintendedapplications.
Understandandfollowallthesafetyinstructionscontainedinthemanual.
Failuretofollowtheinstructionsmaycauseinjury,instrumentdamageand/
ordamageto equipment under test. Kyoritsu is by no means liableforany
damageresultingfromtheinstrumentincontradictiontothiscautionarynote.
Thesymbol
#
indicatedontheinstrumentmeansthattheusermustrefer
torelatedpartsofthemanualforsafeoperationoftheinstrument.Besure
tocarefullyreadtheinstructionsfollowingeach
#
symbolinthismanual.
#
DANGER
is reservedforconditionsandactionsthatare likelyto
causeseriousorfatalinjury.
#
WARNING
is reservedforconditionsandactionsthatcancause
seriousorfatalInjury.
#
CAUTION
isreservedforconditions and actions that can cause
minorinjuryorInstrumentdamage.
Followingsymbolsareused on the instrument and in theinstruction
manual.Attentionshouldbepaidtoeachsymboltoensureyoursafety.
#
Refertotheinstructionsinthemanual.
Indicatesaninstrumentwithdoubleorreinforcedinsulation.
Indicatesthatthisinstrumentcanclamponbareconductorswhen
measuringavoltagecorrespondingtotheapplicableMeasurement
category,whichismarkednexttothissymbol.
IndicatesAC(AlternatingCurrent).
IndicatesEarth.
#
DANGER
Nevermakemeasurementona circuit having potential of300VACor
greater.
Do not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking,
whichleadstoanexplosion.
The transformer jaws are made of metal and their tips are not
completelyinsulated.Beespeciallycarefulaboutthepossibleshorting
wheretheequipmentundertesthasexposedmetalparts.
Neverattempttousetheinstrumentifitssurfaceoryourhandiswet.
Donotexceedthemaximumallowableinputofanymeasurementrange.
Neveropenthebatterycompartmentcoverwhenmakingmeasurement.
Never try to make measurement if any abnormal conditions, such as
brokenTransformerjawsorcaseisnoted.
Theinstrumentistobeusedonlyinitsintendedapplicationsorconditions.
Otherwise,safetyfunctionsequippedwiththeinstrumentdoesn'twork,and
instrumentdamageorseriouspersonalinjurymaybecaused.
Keepyourfingersandhandsbehindthebarrierduringameasurement.
#
WARNING
Never attempt to make any measurement, if any abnormal conditions
arenoted,suchasbrokencase,crackedtestleadsandexposedmetal
parts.
Do not install substitute parts or make any modification to the
instrument. Return the instrument to Kyoritsu or your distributor for
repairorre-calibration.
#
CAUTION
Make sure that the range selector switch is set to an appropriate
positionbeforemakingmeasurement.
Donotexposethe instrument to thedirectsun,extremetemperatures
ordewfall.
Be sure to set the range selector switch to the "OFF" position after
use.Whenthe instrumentwillnotbe inusefora longperiodoftime,
placeitinstorageafterremovingthebatteries.
Use a damp cloth and detergent for cleaning the instrument. Do not
useabrasivesorsolvents.
Always switch off the instrument before opening the battery
compartmentcoverforbatteryreplacement.
Measurementcategories(Over-voltagecategories)
Toensuresafeoperationofmeasuringinstruments,IEC61010establishessafety
standards for various electrical environments, categorized as O to CAT IV, and
calledmeasurementcategories.
Higher-numbered categories correspond to electrical environments with greater
momentary energy, so a measuring instrument designed for CAT III environments
canenduregreatermomentaryenergythanonedesignedforCATII.
O: Circuitswhicharenotdirectlyconnectedtothemainspowersupply.
CATII: Primary electrical circuits ofequipmentconnectedto an AC electrical
outletbyapowercord.
CATIII: Primary electrical circuits of the equipment connected directly to the
distributionpanel,andfeedersfromthedistributionpaneltooutlets.
CATIV: The circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution
panel).
2. FEATURES
DigitalclamptesterforACleakagecurrentmeasurement.
Least affected by external magnetic field, providing wide measuring
rangefromverysmalltolargecurrents.
Designed to safety standard IEC 61010-2-032: measurement CAT.
,
300Vandpollutiondegree2.
Data hold function to allow for easy readings in dimly lit or hard-to-
reachlocations.
Providesfilteringfunctiontoremovehighfrequencygeneratedbysuch
equipmentasinverters.
Sleepfunctionpreventsunnecessarypowerconsumption.
Dynamicrangeof4000countsfullscale.
Largeeasy-to-readLCDdisplaywithletterheightof13mm.
Transformerjawsfittedwithguardtofurtherimprovesafety.
O: Device which is
not directly
connected to the
mains power supply
3. SPECIFICATIONS
Measuringrangesandaccuracy
Range MeasuringRange Accuracy(Frequencyrange)
400mA 0
399.9mA
 ±
2.0%rdg
±
4dgt
50/60Hz
 ±
3.0%rdg
±
5dgt
40
400Hz
4A 0
3.999A
100A 0
100.0A
When measuring current which pulse element is superposed, differences of the
indicatedvaluemaybecaused betweenranges, ifthepeakvalueexceedsthe
measurement range to a large extent. In this case, the reading at the bigger
rangeshouldbetakenasarightvalue.
Sensitive transformer jaws are used for Leakage clamp meter.Because of the
characteristics of transformer jaws, which can be opened and closed, it is
impossibletoeliminatetheinterferenceofexternalmagneticfieldcompletely.If
there are something, which generating large magnetic field, at a nearby site,
current valuecan be displayed (
0
cannot be displayed.)before clampingon
theconductor.Forsuchacase,pleaseusetheinstrumentatalocationfarfrom
thething,whichgeneratingmagneticfield.
Followingarethetypicalthingsgeneratingmagneticfield.
Conductor fed large current, Motor, Equipment which has magnet, Integrating
wattmeter
Measuring method:
DualIntegration
Display:
 Liquid crystal display with maximum reading of
3999
Low battery warning:
"BATT"markappearsonthedisplay
Overrange Indication:
"OL"appearsonthedisplay whenupperlimit of
measuringrangeisexceeded
Response Time:
Approx.2seconds
Sample Rate:
Approx.2.5timespersecond
Location for use
Indooruse,Altitudeupto2000m
Accuracy-insured
2 3
℃±
5
,relativehumidity85%orless
Temperature and
(withoutcondensation)
Humidity Ranges:
Operating Temperature
0-40
,relativehumidity85%orless
and Humidity Ranges:
(withoutcondensation)
Storage Temperature
-20-60
,relativehumidity85%orless
and Humidity Ranges:
(withoutcondensation)
Power Source:
Two1.5VR03(UM-4)batteries
Current Consumption:
Approx.4mA
Measurement Time:
Approx.150hours
Sleep Function:
Automatically powered down in about 10
minutesafterthelastswitchoperation
Safety Standard:
IEC61010-1,IEC61010-2-032
measurement CAT.
300V, pollution degree 2
IEC61326(EMC),EN50581(RoHS)
Overload Protection:
120Armsmax.for10seconds
Withstand Voltage:
3470Vrms (50/60Hz) for 5 seconds between
metalpartoftransformerjawsandhousingcase
(excepttransformerjawcase)
Insulation Resistance:
10M
Ω
orgreaterat 1000V between metal part
of transformer jaws and housing case (except
transformerjawcase)
Conductor Size:
Approx.28mmindiametermax.
Dimensions:
 169(L)
×
75(W)
×
40(D)mm
Weight:
Approx.220gincludingbatteries
Accessories:
TwoR03(UM-4)batteries
CarryingcaseModel9097
Instructionmanual
Optional Accessories:
Multi-TranModel8008
4. INSTRUMENT LAYOUT
5. PREPARATIONS FOR MEASUREMENT
5-1 Checking Battery Voltage
Set the Range Selector Switch to any position other than the OFF
position. If the marks on the display is clearly visible without "BATT"
mark showing, battery voltage is OK. If the display blanks or "BATT" is
indicated, replace the batteries according to section 8: Battery
Replacement.
NOTE
When the instrument is left powered on, the Sleep function
automaticallyshut the power off; The display blanks even if the
Range Selector Switch is set to a position other than the OFF
positioninthisstate.Topowerontheinstrument,turntheRange
SelectorSwitchorpressanybutton.Ifthedisplaystillblanks,the
batteriesarecompletelyexhausted.Replacethebatteries.
5-2 Checking Switch Setting
Make sure that the Range Selector Switch is set to the appropriate
range.
Alsomakesurethatdataholdfunctionisnotenabled.Ifinappropriate
rangeisselected,desiredmeasurementcannotbemade.
Transform Jaws
Trigger
Range Selector Switch
Hand Strap
Display
Frequency Selector Button
Data Hold Button
●LCD
Low Battery Warning
Frequency Response :
Wide
Frequency Response :
50/60Hz
Data Hold Indication
Unit
Barrier
Barrier is a part providing protection against electrical shock and
ensuring the minimum required air and creepage distances.


Specyfikacje produktu

Marka: Kyoritsu
Kategoria: multimetr
Model: 2434

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Kyoritsu 2434, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje multimetr Kyoritsu

Kyoritsu

Kyoritsu KEW1021R Instrukcja

14 Stycznia 2025
Kyoritsu

Kyoritsu KEW 2510 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KEW 2500 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KEW 2413R Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KT 200 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu 2431 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu 2434 Instrukcja

30 Września 2024

Instrukcje multimetr

Najnowsze instrukcje dla multimetr