INSTRUCTION MANUAL 
DIGITAL MULTIMETER 
KEW1021R 
1. Safety Warnings 
             This  instrument  has  been  designed,  manufactured  and  tested 
according to IEC 61010: Safety requirements for electrical equipment, 
and  delivered in the best  condition  after passing  quality  control tests. 
This instruction manual contains warnings and safety rules which have 
to be observed by the user to ensure safe operation of the instrument 
                   and  to  maintain  it  in  safe  condition.  Therefore,  read  through  these 
operating instructions before using the instrument. 
WARNING 
● Read through and  understand  the instructions  contained in this 
manual before using the instrument. 
 ● Keep  the manual  at hand  to  enable quick reference  whenever 
necessary. 
● The instrument is to be used only in its intended applications. 
● Understand and follow all the safety instructions contained in the 
manual. 
               Failure  to  follow  the  instructions  may  cause  injury,  instrument 
damage and/or damage to equipment under test. Kyoritsu is by no 
                 means  liable  for  any  damage  resulting  from  the  instrument  in 
contradiction to this cautionary note. 
                       The  symbol      indicated  on  the  instrument  means  that  the  user  
must refer to the related parts in the manual for safe operation of the 
                 instrument.  It  is  essential  to  read  the  instructions  wherever  the 
   symbol    appears in the manual.    
 DANGER is reserved for conditions and actions that   
are likely to cause serious or fatal injury. 
 WARNING is reserved for conditions and actions that   
can cause serious or fatal injury. 
 CAUTION is reserved for conditions and actions that   
can cause injury or instrument damage. 
● Symbols listed below are used on this instrument. 
User must refer to the manual. 
   Instrument with double or reinforced insulation. 
       Ground (Earth) AC                             DC
This instrument complies to WEEE Directive (2002/96/EC). 
Please contact your local distributor at disposal. 
Measurement Category 
O Circuits which are not directly connected to the mains power supply. 
  CAT II Primary electrical  circuits  of equipment  connected to an AC 
electrical outlet by a power cord. 
CAT III Primary electrical circuits of the equipment connected directly 
               to  the  distribution  panel,  and  feeders  from  the  distribution 
panel to outlets. 
  CAT IV The circuit from the service drop to the service entrance, and   
 to the power meter and primary over current protection   
device(distribution panel). 
DANGER 
● Never make measurements under the circumstances exceeding the 
               designed  measurement  category  and  the  rated  voltage  of  the 
instrument. 
● Do not attempt to make measurement in the presence of ammable 
gasses. Otherwise,  the  use of the instrument  may  cause  sparking, 
which can lead to an explosion. 
● Never attempt to use the instrument if its surface or your hand is wet. 
                   ●  Do  not  exceed  the  maximum  allowable  input  of  any  measuring 
range. 
● Never open the Battery compartment cover during a measurement. 
● To avoid electrical shock by touching the equipment under test or its 
surroundings, be sure to wear insulated protective gear. 
● Test  leads  to  be  used for voltage measurements  shall  be  rated  as 
                 appropriate  for  Measurement  Category  III  or  IV  according  to  IEC 
61010-031 and shall have a voltage rating of 600V or higher. 
● Protective  ngerguard  on the test  leads provide  protection to keep 
                 your  ngers  and  hands  from  touching  an  object  under  test.  Keep 
               your  ngers  and  hands  behind  the  protective  ngerguard  during 
measurement. 
WARNING 
● Never attempt to make measurement if any abnormal conditions, 
such as broken case and exposed metal parts are found on the 
instrument or test leads. 
● Verify proper operation on a known source before use or take action 
as a result of the indication of the instrument. 
● Firmly attach the caps to the test leads when performing 
measurements in CAT III or higher test environments. When 
KEW1021R and the test leads are combined and used together, 
whichever is lower category & voltage to earth either of them belong 
to is applied. 
 ● Do not rotate the Function Switch if the instrument and the   
equipment under test are connected. 
 ● Do not install substitute parts or make any modication to the   
 instrument. For repair or re-calibration, return the instrument to your   
local KYORITSU distributor. 
●Stop using the test lead if the outer jacket is damaged and  the inner 
metal or color jacket is exposed. 
CAUTION 
● Use  of this  instrument is  limited  to domestic,  commercial  and light 
industry  applications.  Strong  electromagnetic  interference  or strong 
magnetic elds, generated by large currents, may cause malfunction 
of the instrument. 
● Firmly insert the test leads. 
● Do not pull or twist the test leads to prevent the risk of damage. 
                     ●  Power  o  the  instrument  after  use.  Remove  the  battery  if  the 
instrument is to be stored and will not be in use for a long period. 
       ●  Do  not  expose  the  instrument to  direct  sunlight,  high  temperature 
and humidity or dewfall. 
               ●  Use  a  cloth  dipped  in  water  or  neutral  detergent  for  cleaning  the 
instrument. Do not use abrasives or solvents. 
NOTE 
● The LCD shows some digits at the Voltage   the C rent  nge even or ur ra
while the test leads are open. And, it may show some digits instead 
of 0 even if the test leads are shorted.  However, these phenomena 
don’t affect measurement results. 
● A resistance measurement takes time to settle the reading if there is 
a high resistance or capacitance components. 
2. Specication 
● Accuracy (Temperature: 23 ± 5°C, Humidity: 45 -  ) 75%
                      ACV RMS    /                                                            
             (Auto Range) 
Range 
Display Range 
Accuracy (sine wave) 
6V 
0.000, 0.006 - 6.299V 
 ±1.0 %rdg±3dgt   
( -500Hz) 40
60V 
5.70 - 62.99V 
600V 
57.0 - 629.9V 
   Guaranteed accuracy : 0.01V-600V,    less than 900V peak 
Input impedance : approx.   10MΩ
              Hz    Frequency -           ACV measurement   
          (Auto Range)
Range 
Display Range 
 Accuracy   
(sine wave) 
99.99Hz 
1 00 - 99.99  0. Hz
± 1%rdg±3dgt 0.
999.9Hz 
95.0 - 999.9Hz 
9.999kHz 
0.950 - 9.9 kHz 99
99.99kHz 
95.0 - 99.99kHz 
Guaranteed accuracy : 10 - kHz Hz 99
    %    DUTY -   ACV measurement
Range 
Display Range 
 Accuracy   
(Square wave) 
99.9 % 
0.0   99.9 % –
±1.0%rdg±3dgt 
   (50/60Hz) 
Guaranteed accuracy : 10%-90% 
                          DCV                                                   (Auto Range) 
Range 
Display Range 
Accuracy 
6.000V 
0.000 - ±6.299V 
±0.5 %rdg±3dgt 
60.00V 
±5.  - ± . V 70 62 99
600.0V 
±  - ± 9.9V 57.0 62
Guaranteed accuracy : 0V-±600V 
Input impedance : approx. 11M (6V range) /  (60/600V range) Ω10MΩ
  ACmV / RMS    
Range 
Display Range 
Accuracy (sine wave) 
600.0mV 
0.0, 0.  9 - 629.9mV 
 ±2.0 %rdg±3dgt   
(40-500Hz) 
   Guaranteed accuracy  1:  .2mV-600mV,    less than 900mV peak 
Input impedance : approx. 900kΩ 
                  AC Clamp Sensor    / RMS                                (Auto Range) 
Range 
Display Range 
Accuracy (sine wave) 
60.00A 
0.00, 0.09 - 62.99A 
±2.0 %rdg±3dgt 
   + Sensor accuracy (40-500Hz) 
200.0A 
57.0 - 209.  9A
Direct reading from 10mV / A output Clamp sensor 
   Guaranteed accuracy  0.12A-200A,    less than 300A peak : 
Input impedance : approx. 900kΩ 
DCmV 
Range 
Display Range 
Accuracy 
600.0mV 
0.0 - ±629.9mV 
±1.5 %rdg±3dgt 
   Guaranteed accuracy : 0mV-±600mV,    Input impedance : approx. 900k  Ω
                                DC Clamp Sensor                                            (Auto Range) 
Range 
Display Range 
Accuracy 
60.00A 
0.00 - ±62.99A 
±1.5 %rdg±3dgt 
200.0A 
±57.0 - ±209.9A 
+ Sensor accuracy  
Direct reading from 10mV / A output Clamp sensor 
   Guaranteed accuracy : 0A-±2 A,    Input impedance : approx. 900k  00 Ω
            Resistance                            (Auto Range) 
Range 
Display Range 
Accuracy 
600.0  Ω
0.0 - 629.9  Ω
±0.5 %rdg±4dgt 
6.000k  Ω
0.570 - 6.299k  Ω
±0.5 %rdg±2dgt 
60.00k  Ω
5.70 - 62.99k  Ω
600.0k  Ω
57.0 - 629.9k  Ω
6.000M  Ω
0.570 - 6.299M  Ω
40.00M  Ω
5.70 - 41.99M  Ω
±1.5 %rdg±3dgt 
   Guaranteed accuracy : 0 - M   Open-loop Voltage : less than   Ω40 Ω,  3V
Measurement current : less than 1mA 
 Continuity 
Range 
Display Range 
Accuracy 
600.0  Ω
0.0 - 629.9  Ω
Buzzer threshold value : 
less than   90Ω
Open-loop Voltage : less than 3V, Measurement current : less than 1mA 
 Diode 
Range 
Display Range 
Accuracy 
2.000V 
0.000 - 2.09  9V
±5 %rdg±5dgt 
   Guaranteed accuracy : 0V-2V,    Open-loop Voltage : less than 3V 
Measurement current : approx.0.5mA (Vf=0.6V) 
           Capacitance                               (Auto Range) 
Range 
Display Range 
Accuracy 
60.00nF 
0.00 - 62.99  nF
±2.0 %rdg±5dgt* 
600.0nF 
57.0 - 629.9nF 
6.000µF 
0.570 - 6.299µF 
±5.0 %rdg±5dgt 
60.00µF 
5.70 - 62.99µF 
600.0µF 
57.0 - 629.9µF 
1000µF 
570 - 1049µF 
* Accuracy after canceling oating capacitance using REL function. 
Guaranteed accuracy : 0 -1000µFnF  
 ACA / RMS                                   (Auto Range) 
Range 
Display Range 
Accuracy (sine wave) 
6.000A 
0.000, 0.006 - 6.299A 
±1.5 %rdg±3dgt 
(40-500Hz) 
10.00A 
5.  - 10.49A 70
 Guaranteed accuracy : 0.01 - 10A, less than 15Apeak   A 
         Hz    Frequency - ACA measurement                    (Auto Range) 
Range 
Display Range 
Accuracy 
99.99Hz 
10.00 - 99.99Hz 
±0.1 %rdg±3dgt 
999.9Hz 
95.0 - 999.9Hz 
9.999kHz 
0.950 - 9.999kHz 
Guaranteed accuracy : 10Hz-9.9kHz 
    %    DUTY - ACA measurement 
Range 
Display Range 
Accuracy 
99.9 % 
0.0   99.9 % –
±1.0%rdg±3dgt (50/60Hz) 
Guaranteed accuracy : 10%-90% (Square wave) 
                       DCA                                                                (Auto Range) 
Range 
Display Range 
Accuracy (sine wave) 
6.000A 
0.000 - ±6.299A 
±1.5 %rdg±3dgt 
10.00A 
±5.70 - ±10.49A 
Guaranteed accuracy : 0A-±10A 
 ●Measuring method :  ⊿Σ modulation 
●Over-range indication : OL 
●Measurement cycle : 2.5 times per second 
(1000µF range of Capacitance function 0.  times per second) 05
●Crest factor : less than 3 (45-65Hz) 
For non-sinusoidal waveforms, add ±0.5 %rdg±5dgt 
(Applicable functions : ACV, ACmV, AC clamp sensor, ACA) 
●Applicable standards : 
IEC 61010-1 /61010-2-033 : CAT IV 300V  CAT III 600V / 
Pollution degree 2, Indoor use, Altitude up to 2000m 
IEC61010-31 (Test leads Model 7066A) 
IEC 61326 (EMC) , EN 50581 (RoHS) 
●Withstand voltage : AC5160Vrms 5sec between circuit and enclosure 
●IP rating  IP40 (IEC60529) : 
 ●Insulation resistance : 100MΩ or more /1000V between enclosure   
and electrical circuit 
●Operating temperature and humidity range : 
0 to °C, 80%RH or less (no condensation)  40
●Storage Temperature and humidity range : 
-20 to 60°C, 80%RH or less (no condensation) 
●Power source : DC3V R03/LR03 (AAA) × 2 
●Current consumption : 3mA or less 
●Battery life : Approx. 200 hours (ACV, continuous, no load, with R03) 
 ●Dimension, Weight : 155(L)× (W)× (D)mm  75 40 , 
approx. 250g (including batteries and Wing-type holder) 
●Accessories : Test leads (M-7066A), soft case(M-9097) 
Instruction manual 10A/600V Fuse (M-8919 included) ,  , 
Function Switch 
Cap 
Protect ngerguard 
REL / MAX MIN Key ・
SELECT Key 
Voltage Terminal 
COM Terminal 
HOLD / Back light Key 
LCD 
Current Terminal 
Protective gerguard: 
It is a part providing protection against electrical shock and ensuring the 
minimum required air and creepage distances. 
Cap: 
                           Test  leads  can  be  used  under  the  CAT  II  and  CAT  III  and  CAT  IV 
environments by  attaching a protective  cap as illustrated below. Use of 
                 our  protective  cap  offers  dierent  lengths  suitable  for  the  test 
environments. 
                   When  the  instrument  and  the  test  lead  are  combined  and  used 
together, whichever lower category either of them belongs to will be 
applied. 
This instrument is designed for CAT IV 300V/ CAT III 600V. Test leads 
M-7066A with the supplied caps are designed for CAT IV600V/ CAT 
III 1000V and without the caps are for CAT II 1000V. 
Specyfikacje produktu
| Marka: | Kyoritsu | 
| Kategoria: | multimetr | 
| Model: | KEW1021R | 
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Kyoritsu KEW1021R, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje multimetr Kyoritsu
                        
                         14 Stycznia 2025
                        
                    
                                                            
                        
                         14 Grudnia 2024
                        
                    
                                                            
                        
                         14 Grudnia 2024
                        
                    
                                                            
                        
                         14 Grudnia 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            
                        
                         30 Września 2024
                        
                    
                                                            Instrukcje multimetr
- Keithley
- Kingcraft
- Milwaukee
- Schneider
- IWH
- Metrix
- HT Instruments
- BaseTech
- Beha-Amprobe
- Greenlee
- Altai
- Testec
- Ferm
- Appa
- Megger
Najnowsze instrukcje dla multimetr
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         3 Kwietnia 2025
                        
                    
                                                            
                        
                         28 Marca 2025
                        
                    
                                                            
                        
                         26 Marca 2025
                        
                    
                                                            
                        
                         14 Marca 2025