用 户 手 册
Users Guide
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
RP1010H高压示波器探头
High Voltage Probe for Oscilloscope
北京普源精电科技有限公司 RIGOL Technologies,Inc.
邮箱 :网址E-mailservice@rigol.com Websitewww.rigol.com
Warning
Do not attempt to take measurements from sources when
the chassis or return lead is not ground.
This ground connection is critical to the safety operation of
the probe. Failure to make this connection may result in
personal injuries or damages to the probe or voltmeter. This
connection must be made before the probe tip comes into
contact with the high voltage and must not be removed
until after the probe tip has been removed.
Do not connect the ground clip to the high voltage source or
the probe tip to the ground for any reason.
BEFORE turning the high voltage on, make sure that no
part of your body is in contact with the device.
Remembering that the voltage being measured is 1000
times greater than the voltmeter reading.
When the measuring voltage are DC 8 to 10 KV, or
AC rms 5 to 7 KV or AC peak 15 to 20 KV, do not
exceed 60 sec on each measurement. You can break off for
5 minutes at least during the measurement. Otherwise
overheating will affect the measurement accuracy even
damage the probe. Disconnect the probe tip from the high
voltage source removing the ground clip lead. BEFORE
Clean only the exterior probe body and cables. Use a soft
cotton cloth lightly moistened with a mild solution of
detergent and water. Do not submerge any portion of the
probe at any time.
Dry the probe thoroughly before any voltage measurement.
Make sure the probe is free from any solvents or solvent
fumes as these can case deterioration to the probe body and
cables.
When the measuring voltage is under
the followings. Each measurement
should be within 60 sec
.
You can break
off for 5 minutes at least during the
measurement.
DC: 8 ~ 10 KV
AC RMS: 5 ~ 7 KV
AC PEAK: 15~ 20 KV
当测量电压为下述情况时,每
次测量时间不得超过 期60秒。
间需断开并等待散热至少5分
钟。
直流:8 ~ 10 KV
交流 有效值:5 ~ 7 KV
交流 峰峰值:15 ~ 20 KV
!
Cleaning
性能指标 (Specifications
最大工
作电压Max. Working Voltage
DC: 0 ~ 10 KV DC
AC: pulse 20 KV peak to peak
AC: sine wave 7KV rms
信噪比Signal/Noise DC 60 dB (1 KHz),
50 dB (1 MHz)
带宽Bandwidth DC ~40 MHz (-3 dB)
衰减比例Division Ratio 1:1000
上升时间Rise Time 8.8 ns
输入阻抗Input Impedance 100 M 5% Ω±
输入电容Input Capacitance 3.0 pF ±0.5 pF
补偿范围Compensation Range 5 pF ~ 50 p
F
线长度Cable Length 2.0 meter ( 0.2 M) ±
温度系数Temperature Coefficient ≤ ℃200 PPM /
精确度Accuracy
DC
±3% (DC to
10 KV)
AC
±3% (1 KHz/1 KV )
-3dB 40 MHz
安全认证Safety App. CE
操作温度Operating Temperature 0 ~ 50
存储温度Storage Temperature -20 ~ +70
重量 尺寸/ Weight / Dimension 250 g /
Φ
75 x 340 L
mm
SM10 2 0 0 0 2 5 7 9

Oceń tę instrukcję

4.3/5 (6 Recenzje)

Specyfikacje produktu

Marka: Rigol
Kategoria: Niesklasyfikowane
Model: RP1010H

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Rigol RP1010H, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Rigol

Rigol

Rigol NFP-3 Instrukcja

8 Października 2024
Rigol

Rigol PVA7250 Instrukcja

7 Października 2024
Rigol

Rigol PHA1150 Instrukcja

7 Października 2024
Rigol

Rigol RM-DG Instrukcja

30 Września 2024
Rigol
Rigol

Rigol PVP2150 Instrukcja

30 Września 2024
Rigol

Rigol PVP3150 Instrukcja

29 Września 2024
Rigol

Rigol PVP2350 Instrukcja

29 Września 2024
Rigol

Rigol PVA8700 Instrukcja

29 Września 2024
Rigol

Rigol RP5600A Instrukcja

29 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

Smallrig
Qsc

QSC TSC-8 Instrukcja

29 Stycznia 2025
Xtrarm
Sunding
Xtrarm
Xtrarm
Osee

OSEE TX402 Instrukcja

29 Stycznia 2025