Tester for systems check
INSTRUMENT CHARACTERISTICS AND FUNCTIONS
Voltage, frequency and current measurement
Voltage between L and N
Voltage between L and PE, N and PE, L and N
Voltage between probe and PE
Dispersed current measurement up to 1 A with clamp meter
Operating current measurement up to 150 A with clamp meter
Dierential protective devices check
Measurement of contact voltage without intervention
Intervention time measurement
Intervention test with nominal fault current
Tests on circuits and dierential switches with increasing fault current
 Testsondierentialswitcheswith5•I∆N (10 mA and 30 mA)
Tests on dierential switches suitable for direct pulsating fault currents
Tests on special dierential switches
Test with adjustable fault current
Test of the dierential protection devices in IT systems
Magnetothermal switches check
Measurement of the impedance of the fault ring and determination
of the short circuit current up to 50 KA
Determination and summary of the suitable magnetothermal switches
Meters start test
Circuit impedance measurement
Earth resistance measurement
Floor and insulating walls impedance measurement
Insulating resistance measurement
Check of the continuity of the earthing and equipotential protection leads
Phase sequence recording
Code Model Description
VP330500 Multitest VE 100 SII Universal measuring device to check the systems
Universal instrument to check electrical systems in compliance with
the CEI 64-8 standard (LAW 46-90).
Built in compliance with the strictest European standards, it is
equipped with a large display to view the connection diagrams of
each of the selected tests.
By pressing a key once again it also shows the information neces-
sary to make correct measurements (instructions integrated in the
instrument).
All the tests are managed is a simple and straightforward manner
(main functions/subfunctions) by means of two keys.
This allows the measurements requested by the law in force to be
taken for a reliable and fast checking of electric systems.
The Multistamp VE 110 II accessory module enables the storage and
printing of the data recorded and its transfer to a PC, if necessary, to
widen its characteristics further.
Multitest VE 100 SII
Ease of use:
3 keys to control the instrument:
“MENU” key: to select basic functions and subfunctions
“START” key: to start the measurement cycle
“ I∆N/i” key: for dierential switch intervention
and to access the “ON-LINE” guide
Printer, memory and interface
Backlit display
9 positions function selector
Adjustable reading angle
Socket for battery charger and clamp meter
Automatic settings, surveillance and switching o
ELECTRIC SYSTEMS CHECK
120/230/400 V nominal voltage systems
16/50/60/200/400 Hz nominal frequency systems
User guide instructions integrated in the instrument
Power supply: 6 1.5 V stilo batteries (IEC LR6) or NiCd or NiMH accumulators
3 digits backlit LCD display (70x37 mm) with 128x64 pixel matrix
Operating temperature: -10 °C ÷ +50 °C
Storage temperature: -20 °C ÷ +60 °C
Checks:
- Voltage measurements UL-N
- Voltage measurements UL-PE / UL-L
Concatenated voltage, phases cycle
Voltage measurement between PE and probe
- Frequency measurements
- Continuity of the equipotentials (RL0) polarity automatic inversion
RL0 (polarity + a PE)
RL0 (polarity – a PE)
- Insulation resistance (RISO) 500 V / 250 V / 100 V
- Earth resistance (RE) with and without peg
- Socket measurement without RCD intervention
- Measurement in autorange
- Measurement range: 10 kΩ / 1 kΩ / 100 Ω / 10 Ω
- Dierential test (RCD): 10/30/100/300/500 mA
- Without RCD intervention:
Internal line resistance
Fault ring resistance
- Degree of protection: IP40 instrument, IP40 test point
- Weight: ca. 2.5 kg (with batteries)
- Included accessories
Instructions for use, shoulder strap, case, 2 pole measurement adapter,
3 pole jumper for transformation into adapter, schuko probe adapter,
2 alligator clips, 1 battery set, fuses
REFERENCE STANDARDS
Compliance with Community Directives: 2006/95/EC (Low Voltage) and 2004/108/EC (E.M.C.)
isdeclaredwithreferencetothefollowingstandards:•Safety:EN61010-1•E.M.Compatibility:EN61000-6-4/EN61000-6-1


Specyfikacje produktu

Marka: Vemer
Kategoria: Niesklasyfikowane
Model: Multitest VE 100 SII

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Vemer Multitest VE 100 SII, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Vemer

Vemer

Vemer PR-1DA Instrukcja

4 Października 2024
Vemer

Vemer TCR Instrukcja

3 Października 2024
Vemer

Vemer Set PC RF8 Instrukcja

1 Października 2024
Vemer

Vemer ADR-D Vision Instrukcja

22 Września 2024
Vemer

Vemer Athena.Temp Instrukcja

22 Września 2024
Vemer

Vemer FR NTC-4P3D Instrukcja

21 Września 2024
Vemer

Vemer HT NiPt-2P3A Instrukcja

21 Września 2024
Vemer

Vemer LSM-01 Instrukcja

18 Września 2024
Vemer

Vemer Phono Instrukcja

16 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

SmallRig

SmallRig VB99 SE Instrukcja

29 Stycznia 2025
QSC

QSC TSC-8 Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS5130 Instrukcja

29 Stycznia 2025
Sunding

Sunding SD-581G Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS5070 Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS2010 Instrukcja

29 Stycznia 2025
OSEE

OSEE TX402 Instrukcja

29 Stycznia 2025